Descrição do produto

Electromechanical pressure switch with electrical contact compatible for medical use, class 2.5; suitable for signaling the minimum ramp pressure; 1/4 "M threaded connection for installation on the interception group of medical gas units; degreased for oxygen use, suitable for use with medical gases.

Normativos

EN 837-1 | EN ISO 7396-1 | EN 12094-10 | EN ISO 15001

Componentes

Connection thread: ¼ ”BSP
Body Brass
DOUBLE EPDM gaskets
NO / NC contacts

 

Electrical Features:
Max Load: 0.5 A at 250 Volt AC, 0.15 A at 110 Volt DC
Exchange contacts (Common, NO and NC) - SPDT
Electrical connection according to DIN43650 standards
Max voltage 250VAC - 110VDC
Max current A.C. 0.5A - D.C. 0.15 °
Max pressure 300 bar
Operating temperature: -25 ° C ÷ + 85 ° C
Insulator / Hardener Epoxy resin
Max tightening torque 35 Nm
IP65 protection

Degreased use oxygen

Kit de manutenção

Dados técnicos

CODE THREAD DOWNWARD CALIBRATION GAS
H0171 1/4" G 25 BAR DOWN ALL
H0172 1/4" G 35 BAR DOWN ALL
H0173 1/4" G 100 BAR DOWN ALL

 

 

Imagens